1. The SLED CAM moves to position 2, which moves the SLED PLATE 0.640 cm (0.250 in.)
forward.
2. The DRIVE MOTOR M2 actuates. The DRIVE MOTOR M2 drives the TIMING BELTS,
which rotate the DRIVE ROLLERS. TIMING BELTS are not visible in the graphic.
3. The DRIVE ROLLERS drive the CASSETTE to the back until the CASSETTE REAR
SENSOR S3 detects the CASSETTE.
4. The CASSETTE REAR SENSOR S3 sends a signal to the MSC BOARD A1 to deactuate
the MOTOR.
5. After a delay of 20 ms, the MSC BOARD A1 deactuates the DRIVE MOTOR M2.