2-12 8F2927 January 17, 2007
Using and Maintaining the Imager
2. Open the main door.
3. Pull down on the release knob.
4. Remove the PMM.
5. Install the new PMM.
6. Close the main door.
7. Log the change in the Imager. On the touch screen, touch:
Utilities
Change filter
Log Filter Change
Yes
OK
8. Touch:
Close to return to the Utilities screen, or
Main to return to the Main Screen.