6-4 8F927 January 17, 2007
Specifications
be provided within 2.5 m (8 ft) of the Imager.
The wire must be insulation-rated for 600 V. A dedicated line is 
recommended.
Network Requirements
The 6800 Laser Imager receives digital images from medical imaging devices 
(modalities) over a 10/100 Base T Ethernet Network. A single 10/100 Base T 
Ethernet Network connection via CAT 5 UTP cable terminating in an RJ-45 
plug must be located within 3.05 m (10 ft) of the rear of the Imager.
Film
Prior to installation you must order and have on hand Kodak DryView Laser 
Imaging Film. Contact your sales representative if you have questions.
Film Types The 6800 Laser Imager accommodates the following film types:
• Kodak DryView DVB Laser Imaging Film
• Kodak DryView DVC Laser Imaging Film
• Kodak DryView DVB+ Laser Imaging Film
• Kodak DryView DVB+ Premium Laser Imaging Film
• Kodak DryView Mammography Laser Imaging Film
• Kodak DryView Mammography Plus Laser Imaging Film
NOTE:  Not all film types are available in every country.
Film Sizes The 6800 Laser Imager accommodates the following film sizes:
• For DVB, DVB+, DVB+ Premium, and DVC Laser Imaging Film:
– 20 x 25 cm (8 x 10 in.)
– 25 x 30 cm (10 x 12 in.)
– 28 x 35 cm (11 x 14 in.)
– 35 x 35 cm (14 x 14 in.)
Area Power
North and Central 
America
• 120 V AC +6% to -10%, 60 Hz ±3Hz
• 208 V AC ±10%, 60 Hz ±3Hz
• 240 V AC ±10%, 60 Hz ±3Hz
NOTE:  For 120 V AC installations, a 20 A service is 
required.
Europe 230 V AC ±10%, 50/60 Hz ±3Hz
Japan 100 V AC ±10%, 50/60 Hz ±3Hz