6-2 8F927 January 17, 2007
Specifications
IMPORTANT: 
• This equipment is intended to connect to other medical devices. Only an 
authorized Service Provider of Kodak Products or Customer Qualified 
Service Personnel may install this equipment.
• Do not locate the Imager within 2 m (6.56 ft) of where a cell phone will 
be used, even if the cell phone is separated from the equipment by a 
wall.
• Do not locate the Imager within 4 m (13.12 ft) of a microwave oven. 
Electromagnetic radiation from a microwave oven is only an issue if, 
after the oven door is closed and latched, the seal does not maintain an 
electromagnetic tight fit between the oven door and the oven main 
housing. Determining if the seal has an electromagnetic tight fit requires 
special detection equipment.
• Do not locate the Imager in the presence of flammable anesthetics, 
oxygen, or nitrous oxide. The Imager does not have a gas-sealed 
electronics enclosure and could ignite any flammable or explosive gases 
present in the Imager’s environment.
• Do not locate the Imager in close contact with MRI devices due to 
possible high magnetic fields near an MRI unit. The magnetic field in the 
area where the Imager is installed must be less than 50 Gauss.
• Do not locate the Imager closer than 1.83 m (6 ft) from a patient bed or 
chair.
• Do not substitute or modify any part of the Imager without prior written 
approval of Eastman Kodak Company.
Environmental Requirements
Temperature • Operating: 15° to 35° C (59° to 95° F)
• Storage: -35° to 60° C (-31° to 140° F)
Relative Humidity • Operating: 20% to 85% RH, noncondensing
• Storage: 10% to 90% RH, noncondensing
Altitude • 30 m (100 ft) below sea level to 3,048 m (10,000 ft) above sea level
Floor Vibration For the Imager to operate with no image degradation, floor vibration levels 
cannot exceed the following.
G
2
/Hz Freq. Range Direction
1 x 10
-7
1-50 Hz Vertical or horizontal