EasyManuals Logo

Konica Minolta MULTI GLOSS 268Plus User Manual

Konica Minolta MULTI GLOSS 268Plus
58 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
25
8.2.2 Change cal.values
The gloss values of the calibration standard in the
holder included with delivery are saved in the
measuring device. During automatic calibration, this
data is assigned to the standard in the holder.
In some cases it will be necessary to enter data for
a new calibration standard, for example if the
previous standard has been damaged or scratched.
To ensure exact calibration, only original
standards from the manufacturer should be
used.
You can use the path shown on the right side to
reach the Change cal.values menu option.
At three angle units, a selection menu will appear
for geometries. Select the desired geometry and
press the scroll wheel.
A warning message appears. You can cancel this
process by pressing the operate button.
If you press the scroll wheel, you will continue with
the process of changing calibration values.
In the next display you can enter new calibration
values.
Calibrate

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta MULTI GLOSS 268Plus and is the answer not in the manual?

Konica Minolta MULTI GLOSS 268Plus Specifications

General IconGeneral
BrandKonica Minolta
ModelMULTI GLOSS 268Plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals