Blood Pressure
Blood Pressure
Pressure range Systolic: 0-300 mmHg / Diastolic 0-200 mmHg
Accuracy +/- 4 mmHg
Calibration Pressure sensor must be calibrated to sphygmomanometer.
See Preferences menu.
Pulse
Available pulses Carotid, Brachial, Radial and Umbilical (SimNewB)
Only 3 pulses can be palpated simultaneously. (Power considerations).
Palpated BP Radial pulse turns off at systolic BP.
Brachial pulse turns off at 20 mmHg, to prevent noise generation in auscultation
area.
Adult
Manikin
Child
Manikin
Infant
Manikin
Default Blood Pressures (mmHg) 120/80 100/70 60/40
Adult pulses are automatically adjusted to BP as follows:
Systolic BP Carotid Pulse Radial/Brachial Pulse
>= 88 Normal Normal
< 88 Normal Weak
< 80 Normal Absent
< 70 Weak Absent
< 60 Absent Absent
Infant pulses are automatically set to absent when Systolic BP is below 10 mm Hg.
Otherwise infant pulses are normal.
Adult pulses are automatically adjusted to BP as follows:
Systolic BP Carotid Pulse Radial/Brachial Pulse
>= 88 Normal Normal
< 88 Normal Weak
< 80 Normal Absent
< 70 Weak Absent
< 60 Absent Absent
Heart Sounds
Heart sounds are synchronized to ECG.
Heartsounds Adult Manikin Child Manikin Infant Manikin
Normal
• • •
Aortic Stenosis
• • •
Austin Flint Murmur
• • •
Mitral Valve Prolapse
•
Systolic Murmur
• • •
Diastolic Murmur
•
Friction Rub
•
Opening Snap @70 msec
•
Stills Murmur
• •
Atrial Septal Defect (ASD)
• •
Ventricular Septal Defect (VSD)
• •
Pulmonary Stenosis
•
Lung Sounds
Lung sounds are synchronized to breathing rates, adjustable from 0 – 60 breaths per minute.
Lung Sounds Adult Manikin Child Manikin Infant Manikin
Normal Breath Sounds
• • •
Fine Crackles
• • •
Coarse Crackles
• • •
Pneumonia
• • •
Wheeze
• • •
Stridor
• • •
Pleural Rub
•
Rhonchi
• • •