EasyManua.ls Logo

LDARtools phx42 - Menu Options on phx42 App; Firmware Updates

LDARtools phx42
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
UNCONTROLLED IF PRINTED 14
Fill Adapter maintenance is critical for successful operation of your phx42. See the H
2
Fill Adapter
(LDAR#1260) Maintenance and Installation guide via LDARtools.com → Resources
Documentation.
1. Remove the H
2
quick-fill cover (LDAR#42908).
2. Ensure the following:
(a) Verify no dust or debris is in the H
2
Fill Port.
(b) Verify no dust or debris is in the H
2
Fill Adapter.
(c) Verify the arrow on the red handle is pointing to the
breather side of the fill valve (see photo).
(d) Verify the gauge nearest the cylinder reads 1800
psi. If not, adjust the regulator.
Note: Do NOT exceed 1800 psi or the pressure
relief valve will vent.
(e) If pressure has bled off the H
2
Fill Adapter, turn the red handle 180° for 2 seconds
to bleed any air from the system before connecting to the phx42.
3. Power on the phx42.
4. Connect the H
2
Fill Adapter to the H
2
fill port, then verify the connection by slightly
pulling on the 3-way valve and spinning the collar.
5. Turn the red handle 180° to open valve. The arrow should point toward the fill hose.
6. Wait until the needle on the gauge stops moving.
7. Turn the handle back 180° to close the valve. There will be a slight hiss from the release of
pressure.
8. Pull on the collar of the H
2
Fill Adapter to release it.
9. Twist the H
2
quick fill cover onto the H
2
fill port.
10. Close the bottle. There is no need to bleed pressure off the H
2
fill Adapter.
REMINDER: The phx42 MUST have the H
2
quick-fill cover in place at all times unless filling.

Other manuals for LDARtools phx42