Application Error Codes
If the application you are using with the phx42 gives an error code, this is what your machine is trying
to say:
Too many calibration points.
This application failed to set the date and time. I really like to know at least WHEN I am!
This calibration cannot be deleted. Contact LDARtools Technical Support.
This calibration cannot be possible. I’m reading this signal lower than the last gas you applied.
It has been less than x seconds since the last ignition. Cannot calibrate. Please wait.
I can’t run on H
2
this low!
Negative PPM Value Errors
If a negative PPM value is displayed, it means one of the following:
"Pump at max power for too long, report the issue."
"Probe Flow is blocked. Check Probe/ Change Filter."
"Feels like you just removed my filter. Please give me a fresh filter so we can get back to work."
"Feels like you just removed my filter. Please give me a fresh filter so we can get back to work.
(Filter Calibrated)"
"Looks like we just found a high PPM leak. Wait 30 seconds and reignite me. (Chamber
overtemp)"
"Trying to ignite! If 3
rd
attempt fails, report the issue."