UNCONTROLLED IF PRINTED 27
Application Error Codes
If the application you are using with the phx42 gives an error code, this is what your machine is trying
to say:
Code
Description
5
Too many calibration points.
16
My flame is out.
18
This application failed to set the date and time. I really like to know at least WHEN I am!
19
This calibration cannot be deleted. Contact LDARtools Technical Support.
20
This calibration cannot be possible. I’m reading this signal lower than the last gas you applied.
21
It has been less than x seconds since the last ignition. Cannot calibrate. Please wait.
22
I can’t run on H
2
this low!
Negative PPM Value Errors
If a negative PPM value is displayed, it means one of the following:
Code
Description
-40
"Pump at max power for too long, report the issue."
-41
"Probe Flow is blocked. Check Probe/ Change Filter."
-42
"Feels like you just removed my filter. Please give me a fresh filter so we can get back to work."
-43
"Feels like you just removed my filter. Please give me a fresh filter so we can get back to work.
(Filter Calibrated)"
-45
"Looks like we just found a high PPM leak. Wait 30 seconds and reignite me. (Chamber
overtemp)"
-46
"Trying to ignite! If 3
rd
attempt fails, report the issue."

Other manuals for LDARtools phx42