UNCONTROLLED IF PRINTED 42
Environmental Conditions Affecting Use
The following table describes the conditions where using the device and sample stream monitoring is
recommended:
External
Influence
Reference
Operating
Conditions
Normal
Operating
Conditions
Operative
Limits
Transportation
and Storage
Limits
Ambient
Temperature
23±2 °C
73±2 °F
-20 to +40 °C
-4 to 104 °F
-20 and +40 °C
-4 and 122 °F
-20 to +40 °C
-4 and 122 °F
Ambient
Pressure
860 to 1060
mbar
70 to 108 kPa
20 to 108 kPa
Relative
Humidity
50%±10%
FID: 20 to 95%
noncondensing
15 and 95%
noncondensing
0 to 100%
Conducted
Susceptibility
N/A
(The phx42 is battery-operated)
Conducted
Emission
N/A
(The phx42 is battery-operated)
ESD
Sensitivity
>6000 Volts
Battery
Charging
Voltage
12V (Max.) DC
12V (Max.) DC
N/A
Elevation for
Use
Sea Level
Max 2000M
Max 2000M
Chemical Exposure
The phx42 Gas Analyzer uses a flame ionization detector to analyze atmospheric samples containing
hydrocarbons. Sample gas is collected and passed through a H
2
flame and ions from combustion of the
sample are measured, recorded, and reported. Primary gas used for equipment calibration is Methane
in Air (various concentrations).