Calibration
PHOTON
+
System Guide 26
Connect Cables with PHOTON
+
and Start Software
Cable Connection for Calibration
Connect the BNC cables to all input channels and the OUTPUT channel
at one end. Connect all the other ends together with BNC T ports.
Connect this cable harness to a voltmeter that is certified to the accuracy
standard described above (0.25%).
Power up the PHOTON
+
and wait for at least 10 minutes prior to
continuing. Record the environmental parameters such as pressure,
humidity, etc. if necessary.
Start the Calibration Software
Start the Calibration software by clicking the Calibration icon.
Select New under the File menu:
The user is asked to enter the following information regarding this
calibration: