EasyManua.ls Logo

Leader Electronics Corp. LV 5330 - Configuring User-Defined Correction Tables; Creating User-Defined Correction Tables Using the LV 5330

Default Icon
195 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9. CINELITE Display
9-7
9.6 Configuring User-Defined Correction Tables
The default gamma correction value when measuring f Stop levels is 0.45, but you can also
use a user-defined gamma correction table that matches the gamma characteristics of the
camera that you are using.
There are two types of user-defined correction tables. The first type includes USER-1 to
USER-3 and consists of tables that are created using the LV 5330. The second type includes
USER-A to USER-E and consists of tables that have been created externally using a device
such as a PC. Neither type of user-defined correction table is deleted if you initialize the LV
5330 by following the procedure in section
5.7.1, “Initializing the Settings Using SETUP INIT.”
9.6.1 Creating User-Defined Correction Tables Using the LV 5330
You can create and store up to three user-defined correction tables on the LV 5330.
To create a user-defined correction table, prepare a grayscale chart for which the camera f
Stop values vary by one for each step, and follow the procedure below.
Figure 9-8 Creating user-defined correction tables
1. Press CINELITE.
2. Press F•6 GAMMA, and select USER-1.
In this example, the table for USER-1 is created, but the tables for USER-2 and
USER-3 can also be created in the same way.
3. Press F•7 CAL.
When you press F•7 CAL, a user-defined correction table appears in the bottom left of
the screen, and the luminance (the data dump Y value) appears near the intersection
of the X and Y cursors.
This setting is available when F•6 GAMMA is set to an option other than 0.45.

Table of Contents