28
Instructions for Use V 2.3 RevH – 05/2018
6. Optional Accessories
6.1.3 Fine-directional fixture for specimen
clamps
• Beforethefine-directionalfixtureforspecimen
clamps can be mounted, loosen 4 screws (10)
(Allen key No. 3) and carefully remove the fixture
for specimen clamps from the baseplate (9).
10
11
• Usingthe4suppliedscrews(11) and the Allen
key No. 3, fasten the baseplate to the specimen
head (12).
11
• Now,screwthefine-directionalfixturefor
specimen clamps with the 4 screws (10) and
the Allen key No. 3onto the specimen head.
9
12
If the fine-directional fixture for speci-
men clamps is not used, retain the
baseplate and 4 screws (11) together
with the fine-directional fixture for
specimen clamps!
Fig. 23