35
Leica RM2235
• To mount the requiredinsert(89.1-3) turn
the clamping ring (90) counterclockwise and
remove it.
• Placetherequiredinsertintothetensionring
(90) and screw the tension ring onto the thread
(91) by turning it clockwise.
• Mountthespecimen(67) and clamp it by turn-
ing the clamping ring (90) clockwise.
• Toorienttheinsertedspecimen,insertthepin
(92) into the bore (93) and rotate it counter-
clockwise to release the clamp. You can now
rotate the specimen so that the side you want
faces upwards.
• Tolockitinthepositionyouhavechosen,
tighten the pin (92) by turning it clockwise.
6.2.6 Holder for round specimens
Fig. 31
89.3
67
89.1
93
90
89.2
91
6. Optional Accessories
92
The holder for round specimens is designed to accommodate cylindrical specimens.
Inserts for specimens of 6, 15 and 25 mm diameter are available.