Friedrich Leutert GmbH & Co. KG 17
Digital Pressure Indicator DPI - Bedienungsanleitung
PMI
Connected to USB
Recharging . . .
---} ----------------
3.3 Set the date and time of the handheld device
The correct value of time and date is necessary to ensure an exact temporal
assignment of the measurement results.
The date format is month / day / year, the time is displayed in
hours:Minutes:seconds.
The date / time setting of the handheld device is done in the DPI software
by synchronization with the PC settings, provided the PC is set up correctly.
1) Connect the handheld device to the PC using the USB cable. The
handheld device switches on automatically and the battery charging
process begins.
2) Start the DPI software on the PC.
3) Open the menu [Handheld] and [Sync. Time].
4) Click the button [Synchronize Time to Handheld] button. The date
and time settings of the PC are transferred to in the handheld device
and saved there.
5) If necessary, disconnect the handheld device and PC. The handheld
device switches off automatically.