Status Registers (Group Address = 0x0060)
Address Monitored
Parameter
Type/Unit Description
Digital Output Register at 0x1CA1. Non-resettable
fault.
5= Reserved
4= Reserved
3= Reserved
2= Reserved
1= Reserved
0(LSB)= Reserved
Monitor Values (Group Address = 0x0070)
Address Monitored
Parameter
Type/Unit Description
0x0070 Frequency
Reference
Q15/Hz Nominal speed
0x0071 Motor Frequency Q15/Hz Stator Frequency Applied to the Motor
0x0072 Cumulative
Operation Time at
Previous Fault.
Int/second Time
0x0074 Motor Current Q15/A RMS Current
0x0075 DC Link Voltage Q15/V Voltage
0x0076 Motor Power Q15/W Power
0x0077 Motor
Temperature
Q15/°C Temperature
0x0078 Sink Temperature Q15/°C Temperature
0x0079 Controller, DSP,
Temperature
Q15/°C DSP Temperature
0x007A Vcc Supply Q15/V Rail Voltage
0x007B Rotor Speed Q15/Hz Rotor Speed for IM or Synchronous Machine
Drive Operational History (Group Address = 0x0080)
Address Monitored
Parameter
Type/Unit Description
0x0080 Run Seconds Int/second Drive running time
0x0082 Powered
Seconds
Int/second Drive powered up time
0x0084 Fault Total Count Int/na Number of faults (trips) since manufacture, i.e. since
initialisation of the fault log.
0x0085 Fault Count Int/na Number of faults (trips) since last requested reset of
the fault count.
0x0089 Drive Power
Cycle Count
Int/na Count the number of power-on events.
6.12 Gas ballast operation
Gas ballast is operated by a black plastic knob that may be positioned to
either position 0 (which means the gas ballast is closed), position 1, or any
43
300902516_002 - 10/2020 - © Leybold
Operation