Description
14 300324726_002_C0 - 10/2016 - © Leybold
Technical data for the integrated frequency converter
MAG.DRIVE iM
Supply voltage 200 - 240 V ±10%
Mains frequency 50 / 60 Hz
Power consumption
maximum 750 VA
at ultimate pressure 150 VA
Max. load capability of the relays (interface X1, optional) 32 V, 0.5 A
(for SELV or PELV power circuits only)
Max. load for each valve output 24 VDC, 0.35 A
Temperature during operation 10 - 45 °C
Storage temperature – 10 - 60 °C
Relative air humidity (non condensing) 5 to 85 %
Type of protection IP 54
Overvoltage category II
Contamination level acc. to IEC 61010 2