Converter Control MAG.DRIVE L
51
GA 05.130/3.02 - 05/98
4.5.5 Settings temperature management system
Menu item Description Adjustable value / option Ac-
cess
min. max. stan- Unit
value value dard
Setpoint Temp. Setpoint Temperature 20 75 63 °C r/w
Management System
Warn.Temp.TCU Warning temperature
Temperature Management 0 85 70 °C r/w
System
FailTemp. TCU Shutdown temperature
Temperature Management 0 100 72 °C r/w
System
speaker Intermittent = warning off - r/w
Continuous = failure on -
4.5.6 Settings purge / vent
Menu item Description Adjustable value / option Ac-
cess
min. max. stan- Unit
value value dard
P-Nr. HE Testdis Not used
TCU
Enter
Setpoint Temp
63 °C
Enter
Purge/Vent
Enter
P-Nr. HE-Testdis
0
Enter
r (read) = value can only be read
r /w (read/write) = value can be read and written
Warn. Temp.TCU
70 °C
Enter
Fail Temp. TCU
72 °C
Enter
on
speaker
off
Enter