1.6 Technical data
MAG 1600 C/CT W 1600 C/CT 2000 C/CT W 2000 C/CT
Pumping speed for N
2
measured
with splinter guard (PNEUROP) l·s
-1
1250 1300 1550 1650
Gas flow (continuous operation
with Argon) sccm 1000 1000 1000 1000
Compression for N
2
>10
8
>10
8
>10
8
>10
8
Ultimate pressure as to DIN 28 400 mbar < 10
-8
< 10
-8
< 10
-8
< 10
-8
Max. forevacuum pressure
with Nitrogen mbar 1.5 3.3 1.5 3.3
with Argon mbar 1.5 4.1 1.5 4.1
Rotor Speed min
-1
28,800
Run-up time min < 10
Braking time with/without venting min 1 / < 7.5
Cooling water
Cooling connection,
Swagelok elbow for tube OD mm 6.4 (1/4“)
Cooling water temperature °C 15 to 30
°F 59 to 86
Base flange temperature C version °C 30 to 50
(depending on the load) °F 86 to 122
Base flange temperature CT version °C 30 to 63
(controlled) °F 86 to 145
Weight approx. kg 70 65 68 62
High-vacuum connection flange DN 200 ISO-F 200 ISO-F 250 ISO-F 250 ISO-F
Forevacuum connection flange DN 40 KF 40 KF 40 KF 40 KF
Recommended backing pump DRYVAC 100 50 100 50
Admissible ambient temperature °C 5 to 40
°F 40 to 104
Storage temperature °C -10 to +60
°F 14 to 140
9