8 Maintenance
Procedures
Introduction This chapter provides instructions for routine maintenance that may be
required for the SOLAIR 1.0 CFM particle counter.
The maintenance procedures described in this chapter are not required
on regular or prescribed intervals and should be performed only if the
Service Wrench icon is displayed on the MAIN screen, or if the user
has been requested to do so by Lighthouse Technical Support personnel
or the user has reason to question the data they are receiving from the
instrument.
Safety Before performing any of the maintenance tasks described in this
chapter, read Chapter 1 of this manual and become familiar with the
warnings and caution labels discussed throughout this manual.
Maintenance Calibration
To maintain optimum performance of this instrument, it should be
recalibrated annually by a Lighthouse Authorized Service Provider.
Purge Count Test
This section will provide the user with the procedure to check the
counter for zero counts. A purge filter must be attached to the
instrument and fifteen (15) five (5) minute samples must be taken.
There should be no more than 1 count on average per five-minute
sample.