Lighthouse S3100/S5100/Boulder Counter Operating Manual
4-14 248083389-1 Rev
6. Select Connections and verify that Connect Protocol and
Connection settings are set to the following:
• Protocol: Modbus
• Local Port: 502
• Accept Incoming: Yes.
• Under Modbus Mode, set the Message Timeout to “1000ms” if
not already set.
• Press the OK button.Connections Settings will display Done! as
shown in Figure 4-11.
Figure 4-11 Connections