The HB80 hand control has an optimised ergonomic design shaped for the hand. The hand control is suitable for
all kinds of MEDLINE and CARELINE applications such as hospital beds, patient lifts, treatment and examination
couches etc.
The HB80 hand control is available in versions with up to 10 or 12 activation buttons.
Usage:
Usage temperature: 5 ºC to 40 ºC
Storage temperature: -10 ºC to +50 ºC
Compatibility: Compatible with many LINAK control boxes. For further questions, please ask your local LINAK.
Approvals: IEC60601-1:2005 3
rd
edition, ANSI/AAMI ES60601-1:2005 3
rd
edition, and CAN/CSA-22.2 No 60601-1:2008.
The HB86 version has a shorter distance between the buttons and cannot be approved according to EN IEC60601-2-52
Application Environment 4 for care beds used in Domestic area (or EN1970).
HB80 is designed and tested in accordance with EN60601-2-52 cl. 201.11.6.6.101 (Machine washable medical beds).
The HB80 must hang vertically from it’s hook during the washing process.
In order to maintain the flexibility of the cables, it is important that a coiled cable is placed in such a way that the cable’s
own weight does not strain the coil during the washing process.
Recommendations
Clean the hand control regularly to ensure good hygiene standards.
When a defective HB80 is replaced, check that the new HB80 has exactly the same specification and functionality.
Do not submerge the hand control under water.
Unless otherwise specified or agreed by LINAK - the hand control is only intended to be used on LINAK systems.
When changing hand controls for OpenBus™ systems, the power must be switched off.
It is recommended to check the hand control and cable for damage and holes made by violent handling before washing the application or at least
once a year.
It is recommended to have a parking place for the hand control on the application, where the customer ensures that the hand control does not fall off.
For hand controls with magnets:
• If hand controls with magnet are attached to a smooth surface, a movement or twisting of the cable, for example during transport, can cause the
hand control to move and result in damage if the cable is squeezed.
The force of the magnet depends on the thickness of the lacquering, the lacquering type, stickers, steel thickness etc. It is the responsibility of the
customer to verify that the holding force on the application is acceptable.
It is the responsibility of the user/operator to evaluate any possible risk caused by use of permanent magnets.
Warnings
Do not sit or lie on the hand control. It can cause unintended movement of the application.
There is a risk that items with internal magnet for mounting instead of hook can disturb function of cardiac pacemaker, implantable cardioverter
defibrillators or magnetic implants!
20. HB80 (MEDLINE
®
CARELINE
®
)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Linak COBO20 and is the answer not in the manual?

Linak COBO20 Specifications

General IconGeneral
BrandLinak
ModelCOBO20
CategoryController
LanguageEnglish

Related product manuals