6 Use and Storage Conditions
DANGER!
Danger to life due to electric shock!
To ensure the bed’s class I protection against electric shocks:
Ground the mains.
Use exclusively Hospital Grade or Hospital Only receptacles for grounding.
CAUTION!
Minimal clearance underneath the bed (standard version with 15 cm castors) is 4,4 cm!
Observe the path for any obstacles and avoid collisions and possible damages of any bed´s part on the underca-
rriage.
Do not use bed lifts and hoists for lifting the bed (bed hoists for patients are permitted; clearance for the patient
hoists is 15 cm).
Multicare LE with Symbioso are designed for use in rooms for medical purposes. Electrical installations must therefore meet local
norms laying down the necessary conditions for electrical installations.
Disconnect the bed from the mains in exceptional cases (i.e. lightning or earthquake).
Multicare LE and Symbioso are not suitable for indoor environments:
containing ammable gases (except oxygen cylinders).
NOTE There are applied no deviations to requirements of IEC 60601-1-2 ed. 4
NOTE There are no known other measures for keeping the basic safety based on EMC phenomena.
NOTE Beds equipped with integration module meet standard for IEEE 802.11 b/g/n (2400,0 MHz – 2483,5 MHz, modulation
DSSS (IEEE 802.11 b ), OFDM (IEEE 802.11 g/n) 20MHz bandwidth, EIRP = 0,34 W).

Table of Contents

Other manuals for LINET Multicare LE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET Multicare LE and is the answer not in the manual?

LINET Multicare LE Specifications

General IconGeneral
BrandLINET
ModelMulticare LE
CategoryMedical Equipment
LanguageEnglish

Related product manuals