D9U001MC5-0101_0846
12.1.6 Bed Exit Alarm
Any weight drop of more than 20 kg will activate the Bed Exit Alarm.
To deactivate Bed Exit Alarm:
► Press Bed Exit Button 10 for 2 s.
To activate Bed Exit Alarm:
► Press Bed Exit Button 10 for 2 s.
To switch between Bed Exit Alarm zones:
► Press Bed Exit Button 10 shortly to change zone of the Bed Exit Alarm.
To set up the alarm volume:
► Press button 7 until the desired volume is reached.
NOTE: If the alarm is set to minimum volume the mute mode is on.
Inner Zone monitoring (indicator 8)
► Alarm starts when patient moves from the limited area.
Outer Zone monitoring (indicator 9)
► Alarm starts when patient leaves bed.
NOTE Inner zone alarm is the default mode when the Bed Exit Alarm is activated.
12.1.7 Bed Overload
If the bed load is over 254,5kg:
► Overloading is signalized by long acoustic signal.
► The „Hi“ icon is displayed on the display.
NOTE: If the bed is overloaded then it is impossible to position or manipulate with the bed until the overloading is removed.
NOTE: The bed overloading has always higher priority than Hold Mode and Taring functions.
12.1.8 Bed Underload
In case the bed is underloaded (factory zero – 5kg):
► The display shows icon „Lo“.
12.1.9 Weighing in tilt
The bed can weight in tilt. The guarantee of accuracy is secured by spirit level which is located at the head/foot end of the bed. If
the bubble is in the highlighted circle then the weighting is accurate.
12.1.10 Zeroing Scales
The zeroing can be done only in the range ±5kg from the factory zero. The zeroing is used to reset weight on the display and also
to set up user zero, which sets the maximum weight range of the weighting system.
The zeroing must be done on the empty, unloaded bed without mattress and accessories. The zeroing is done after installation,
weight verication or servicing.
To zero scales:
► Remove all accessories and mattress from the bed. Position the bed about 20 cm above the lowest position and the
mattress platform to the horizontal position. Ensure that nothing touches the bed except you.
► Press button 2 (Zero/T) for 0,5s until weight value starts to ash.
► Press button 2 to conrm zeroing.
„0“ is shown at the display and acoustic signal conrms zeroing.
To cancel zeroing:
► Press button 5 while zeroing.