D9U001MC5-0101_0866
Error/Fault Cause Solution
Adjusting with position buttons not
possible
GO button was not pressed Press the GO button.
Function disabled on Attendant Control Panel Enable disabled function.
Drive motors have no power,
Defective drive motors,
Defective battery
Check the mains connection.
Notify the service department.
Plug inserted incorrectly Insert the mains plug correctly.
Faulty power source. Notify the service department.
Faulty control element Notify the service department.
Faulty mattress platform height/tilt adjus-
tment
There is an object on the undercarriage cover Remove the object.
Function disabled on Attendant Control Panel Enable disabled function.
Drive motors have no power,
Defective drive motors,
Defective battery
Check the mains connection.
Notify the service department.
Plug inserted incorrectly Insert the mains plug correctly.
Faulty power source Notify the service department.
Faulty control element Notify the service department.
Lowering backrest from the upright position
not possible
There is an object under the back- rest or in the
drive mechanism
Remove the object.
Locking handle is defective Notify the service department.
Adjusting siderails not possible The siderail lock is dirty Clean the locking mechanism.
Locking handle is defective Notify the service department.
Faulty brakes Dirt blocking brakes mechanically Clean the brake system.
The brake mechanism is defective Notify the service department.
Mattress not inating SCU mains switch turned off Turn switch on. Green mains
power switch will illuminate.
No power to bedframe Check mains connection.
Mains plug inserted incorrectly Insert mains plug correctly.
Faulty power source Notify the service department
of the manufacturer.
Faulty SCU Notify the service department
of the manufacturer.
Air leaking Check mattress pipe connec-
tion.
CPR valve leaking or open Check CPR valves closed.
Air pipes blocked Check piper are not trapped or
kinked.
Mattress partialy inated Check mattress is unfolded
and at.
Mattress damaged or faulty Notify the service department
of the manufacturer.
Unable to change mode or pressure Go button was not pressed Press the GO button.
Fault symbol illuminated & audio alarm One time exception fault
Persistent reoccuring fault
Switch off SCU & switch back
on, reset to see if failt is self-
-cleared.
Notify service department,
mute audio alarm.
CPR symbol illuminated In CPR mode Press CLP or Max to cancel
17 Troubleshooting
DANGER!
Danger to life due to electric shock!
If a fault occurs ensure the electric motor, power box and other electrical parts checked by qualied personnel
only.