Statuses (Multiboard)
Pop-up Meaning Required Action
LOCK
Function locked. Unlock function if required!
CASS
Incorrectly inserted X-Ray Cassette Holder. Insert X-Ray Cassette Holder correctly!
GO
GO Button not activated. Press GO Button!
SIDE
Lateral Tilt disabled when siderail folded down. Raise siderail up to enable additional Lateral Tilting.
COLL
Antitrendelenburg Tilt and Trendelenburg Tilt disa-
bled during Lateral Tilt.
Undo the Lateral Tilt to continue with Antitrendelen-
burg Tilt or Trendelenburg Tilt.
Positioning blocked to avoid collision of the bed with
oor or collision of the bed with bed equipments.
To continue in positioning, adjust the bed so that
there is no collision.
Horizontal position was reached during tilting.
Press corresponding button to continue in positio-
ning.
15°
Lateral Tilt stopped in 15° when Patient Transfer
Mode is activated.
For information only.
Maximum Lateral tilt was adjusted by foot control
(15°).
For information only.
30°L
Lateral Tilt limited to 15 degrees when Backrest
Angle is in 30 degrees or more.
Lower the Backrest Angle to continue in Lateral
Tilting.
LOAd
Safe Working Load exceeded (more than 10 kg over
Safe Working Load).
Remove load!
Maximum Lateral Tilt 15 degrees (Load more than
150 kg).
Remove load to enable Lateral Tilting again!
Lateral Tilt disabled (Load more than 200 kg). Remove load to enable Lateral Tilting again!
FAIL
System Control Unit disconnected (Symbioso).
Connect to the mains power and connect Symbioso
mattress to System Control Unit.
CPr
The mattress deating failed.
Use manual CPR! (System Control
Unit (Symbioso) is disconnected
or automatic deation is not
available.)
SYS
System Fatal Error.
Contact service department approved by manu-
facturer.
- - - - -
- - - - -
Scale module disconnected and Bed Exit monitoring
disabled.
Contact service department approved by manu-
facturer.
SYSC
Column Unit Error.
Contact service department approved by manu-
facturer.
Hi
Safe Working Load exceeded (from 4,5 kg to 10 kg
over Safe Working Load).
Remove load!
Lo
The bed is underloaded.
Ensure the mattress support platform is not lifted
inconveniently by something and the scales are
appropriately tared.

Table of Contents

Other manuals for LINET Multicare LE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET Multicare LE and is the answer not in the manual?

LINET Multicare LE Specifications

General IconGeneral
BrandLINET
ModelMulticare LE
CategoryMedical Equipment
LanguageEnglish

Related product manuals