pg. 50
■
Select Marker 1 2 3 4
Select one marker among 4 markers. Then, a marker already on, is activated when
selected. When the marker is designated to a different trace while it is already on,
the marker will be inactivated on trace concerned when selected.
MARKER >> Select Marker
■
Normal
When the marker is not yet indicated, a single frequency marker on the currently
selected marker will be activated on the center frequency. When the marker is
already indicated, it becomes activated on a selected location. The marker number
is indicated on the marker. The indications on the activated function block and the
right of the display show the frequency and amplitude of a marker. If you press
Normal, the Delta function goes out, and the activated marker moves to the
location of the Delta marker.
MARKER >> Normal
■
Delta
The second marker gets activated on the location of the first marker.
(
When there is
no marker, two markers appear on the center of the display.) The amplitude and
frequency of the first marker were fixed. The marker number is indicated on the
Delta marker, and the same number is indicated on the standard marker as R.
(
For
instance, 1R
)
The indications on the activated function block and the right of the
display show the difference between the frequency and amplitude of two markers.
MARKER >> Delta
■
Delta Pair Ref Delta
It converts to a mode that allows the two markers can be independently adjusted.
When you press Delta Pair, there is a conversion between Reference marker and
Delta. The reference marker number is indicated on the marker as R, (for instance,
1R) and Delta marker is indicated as a marker number.
MARKER >> Delta Pair