tracer point and the lowest tracer point (or when the span is 0), and the amplitude.
PEAK SEARCH >> PK-PK Search
Continuous Pk On Off
When it is ON, the marker is located on the highest level of trace, and when the
sweep is over, it successively detects the highest level, and locates the marker.
PEAK SEARCH >> Continuous Pk
Mkr CF
It sets the center frequency of the analyzer as the marker frequency. On delta mode,
Mkr CF sets the center frequency as the marker delta value. Mkr CF cannot be
used when it is on zero span.
PEAK SEARCH >> Mkr → CF
5-17. Single Sweep
When the analyzer is on continuous sweep mode, it converts to a single sweep. When the
analyzer is already on single sweep mode, it performs a new sweep once if the trigger meets the
conditions.
SINGLE
SWEEP >> Sweep
5-18. SPAN
It activates the span function, and makes it possible to change the span.
Full Span
It sets the span as the entire frequency scope of the analyzer.
SPAN >> Full Span
Zero Span

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LP Technologies LPT-6000 and is the answer not in the manual?