UltraPulse
0637-129-01, Revision G
56 General Operation
Preoperative Instructions
1 Verify that the laser is properly connected, as instructed in “Connection
Instructions” in this manual.
2 Verify that the delivery system is properly connected, and sterile drape the arm
if necessary, as instructed in the delivery system operator manual.
3 Post the “Laser in Use” warning sign outside the treatment room door.
4 If necessary, turn on the main electrical service (wall circuit breaker).
5 Ensure that all persons in the treatment room are wearing the appropriate laser
safety eyewear. See “Laser Safety Eyewear” in this manual for detailed laser
safety eyewear information.
6 If using the OtoScan, Acuspot, AcuBlade, Multi-Application (Oral
Pharyngeal) handpiece, or the Nasal or Laryngeal Probes, connect the external
Purge Air Compressor, as instructed in “Connection Instructions” in this
manual.
7 Turn on the laser, as instructed in “Laser Console Basics” in this manual.
8 Perform the beam alignment check, as instructed in your delivery system
operator manual.
WARNING - Lasers generate a highly concentrated beam of light that may
cause injury if improperly used. To protect the patient and the operating
personnel, the entire laser and the appropriate delivery system operator
manuals, including all Safety and Regulatory sections, should be carefully read
and comprehended before operation.
WARNING - Before beginning an operative procedure, always verify that the
laser aiming and treatment beams are aligned, as instructed in the delivery
system operator manual. If the aiming and treatment beams are not aligned, do
not use the laser until alignment is corrected by a Lumenis-certified service
technician. Misalignment of aiming and treatment beams may result in laser
exposure to nontarget tissues and possible injury.