EasyManua.ls Logo

MagCore Plus II - Trouble Shooting for Alarm Code

MagCore Plus II
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Content
Causes and solutions
4 DC Motor Up Time Out
[Causes]
The motor is not functional or the block plate of the sensor is at wrong
position.
[Solutions]
(1) Check whether the motor is functional.
(2) Check whether the cable dropped off.
(3) Adjust the position of the block plate of the sensor.
5 DC Motor Down Time Out
[Causes]
The motor is not functional or the block plate of the sensor is at wrong
position.
[Solutions]
(1) Check whether the motor is functional.
(2) Check whether the cable has dropped off.
(3) Adjust the position of the block plate of the sensor.
6 Y_SoftLimt+
[Causes]
Manual Y JOG+ operation exceeds the distance of the software.
[Solutions]
(1) Alarm Reset
(2) Manual Y JOG- operation.
7 Y_SoftLimt-
[Causes]
Manual Y JOG- operation exceeds the distance of the software.
[Solutions]
(1) Alarm Reset
(2) Manual Y JOG- operation.