Portable and mobile RF communications equipment can affect the easyScreen. Install and operate the easyScreen according to the EMC
information presented in this chapter.
The easyScreen has been tested for EMC emissions and immunity as a standalone easyScreen. Do not use the easyScreen adjacent to or stacked
with other electronic equipment. If adjacent or stacked use is necessary, the user should verify normal operation in the configuration.
The use of accessories, transducers and cables other than those specified, with the exception of servicing parts sold by Interacoustics as
replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the device.
Anyone connecting additional equipment is responsible for making sure the system complies with the IEC 60601-1-2 standard.
Guidance and manufacturer’s declaration - electromagnetic emissions
The easyScreen is intended for use in the electromagnetic environment specified below. The customer or the user of the Instrument
should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The easyScreen uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The easyScreen is suitable for use in all commercial, industrial,
business, and residential environments.
Harmonic emissions
IEC 61000-3-2
Complies
Class A Category
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
Recommended separation distances betweenportable and mobile RF communications equipment and the easyScreen.
The easyScreen is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of the easyScreen can help prevent electromagnetic interferences by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the easyScreen as recommended below, according to the maximum output
power of the communications equipment.
Rated Maximum output
power of transmitter
[W]
Separation distance according to frequency of transmitter
[m]
150 kHz to 80 MHz
d = 1.17
𝑃
80 MHz to 800 MHz
d = 1.17
𝑃
800 MHz to 2.5 GHz
d = 2.23
𝑃
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHZ, the higher frequency range applies.
Note 2 These guidelines may not apply to all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Instrument is intended for use in the electromagnetic environment specified below. The customer or the user of the easyScreen
should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test level
Compliance
Electromagnetic environment -
guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
+6 kV contact
+8 kV air
+6 kV contact
+8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be greater than
30%.
Electrical fast
transient/burst
IEC61000-4-4
+2 kV for power supply lines
+1 kV for input/output lines
+2 kV for power supply
lines
+1 kV for input/output lines
Mains power quality should be that of
a typical commercial or residential
environment.
Surge
IEC 61000-4-5
+1 kV differential mode
+2 kV common mode
+1 kV differential mode
+2 kV common mode
Mains power quality should be that of
a typical commercial or residential
environment.
Voltage dips, short
interruptions and voltage
variations on power
supply lines
IEC 61000-4-11
< 5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 sec
< 5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT for 5 sec
Mains power quality should be that of
a typical commercial or residential
environment. If the user of the
easyScreen requires continued
operation during power mains
interruptions, it is recommended that
the easyScreen be powered from an
uninterruptable power supply or its
battery.
Power frequency
(50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
residential environment.
Note: UT is the A.C. mains voltage prior to application of the test level.