EasyManua.ls Logo

Maico easyScreen - Pass-Checker (Optional Use for ABR Hardware Testing)

Maico easyScreen
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Manual easyScreen
8508935 Rev. 4
71
04/04/2019
5.16 Pass-Checker (Optional Use for ABR Hardware Testing)
The Pass-Checker accessory can be optionally used to perform a quality check of the
ABR components of the easyScreen. This can be done on a regular basis or can be
performed if you suspect a problem with the easyScreen hardware. The Pass-Checker
can be purchased from your MAICO representative.
Figure 91
Figure 92
The quality check is performed as follows:
Verify that the electrode lead wires and the insert earphones are connected to
the easyScreen preamplifier cable and that the preamplifier cable is attached to
the easyScreen.
Connect the 3 electrode lead wires to the posts on the Pass-Checker matching
the colors of the wires to the ring around the posts (Figure 91).
Insert the insert earphone adapters securely into the holes on either side of the
Pass-Checker (Figure 92).
If the black EarCup
TM
adapters are used with your system these will fit into the
hole directly.
If the clear eartip adapters are used with your system you will need to place a disposable
eartip onto the adapters before inserting them into the Pass-Checker holes.
Switch on the Pass-Checker by pressing the power button. The amber LED
lights up when the Pass-Checker is turned on.
The Pass-Checker has an auto-off function. If you perform repeated tests during
troubleshooting, make sure that the Pass-Checker is still switched on during
your tests.
Perform a standard binaural (both ears) ABR test with the Pass-Checker attached.
The test should proceed quickly through the impedance and measurement
phases ending in a Pass result for both ears.
See below for symptoms, possible causes and suggested troubleshooting (Table 20).

Related product manuals