Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
6-6
Symptom Cause and Corrective Action
Steps
Portable Will Not Undock
From Docking Station
1. Ensure Portable Lock is not in
2. Latch failure, gently pry latch
down; once removed, replace
latch and/or striker
Watchdog Alarm
1. Other than "Service"
failures
1. Replace Portable System Board
WARNING: IF SYMPTOMS ARE PRESENT AND ARE NOT
CORRECTABLE, DO NOT USE MALFUNTIONING
COMPONENT.