Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
12-1
12 Radical-7 Isolation Block Diagram
FILTER MODULE
Double Insulation 4kV
L1
FERRITE
2.0mH
NOTE: Verify Power Supply Part Number
DC/DC PS
OPTO
100pF
Protective Earth
1000pF
Equipotential Terminal
FUSE 1A
Basic Insulation 1.5kV
680pF
L2
FERRITE
"BF" Type
Applied
Part
680pF
FUSE 1A
20293
0.047uF
DC/DC PS
CF Rated
HANDHELD
Basic Insulation 1.5kV
DOCKING STATION SYSTEM BOARD
12V-SELV
Satshare
Patient Cable
OPTO
AC PLUG
RADICAL ISOLATION BLOCK DIAGRAM
SIP/SOP-Serial Port/Analog Out/Nurse Call
IEC- 60950 Compliant Device
Handheld Interface
POWER SUPPLY
12V-SELV
20241
POWER SUPPLY
Basic Insulation 1.5kV
3.3M
100pF
SW
1000pF