Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Signal Extraction Pulse CO-Oximeter
Monochrome and Color Display
Service Manual

Questions and Answers

  • M
    Monica TerrySep 18, 2025
    What to do if Masimo SET Medical Equipment experiences an SRAM Memory Failure?
    • Y
      Yvonne JamesSep 18, 2025
      If your Masimo SET Medical Equipment experiences an SRAM Memory Failure, replace the Portable System Board.
  • D
    Daisy WilcoxAug 25, 2025
    What to do if the Masimo SET Medical Equipment battery LED is flashing?
    • A
      Abigail NguyenAug 25, 2025
      If the battery LED is flashing on your Masimo SET Medical Equipment, first ensure that Service Mode – Deep Discharge is not enabled. If it is not enabled, allow the NiMH Battery to cool down. Next, verify the fan operation. If the issue persists, replace the NiMH Battery or replace the Docking Station System Board.
  • J
    Joan GeorgeAug 25, 2025
    What to do if the Masimo SET Medical Equipment portable will not power up when undocked?
    • B
      bakerchristopherAug 26, 2025
      If your Masimo SET Medical Equipment portable unit will not power up when undocked, check the NiMH Battery charge. If charging is not successful, replace the battery. If the problem continues, replace the Portable System Board.
  • M
    Mr. Patrick BrownAug 27, 2025
    What to do if the Masimo SET Radical-7 Medical Equipment display does not rotate automatically?
    • N
      Nicole AlvarezAug 27, 2025
      If the display on your Masimo SET Medical Equipment does not rotate automatically, ensure the docking station is powered up. Also, ensure the Power Save feature is not enabled. If problems persist, replace the Docking Station System Board.
  • C
    Carl HensleySep 22, 2025
    What to do if Masimo SET Medical Equipment experiences an Invalid SatShare coefficients?
    • E
      Elizabeth RobertsSep 22, 2025
      If your Masimo SET Medical Equipment experiences invalid SatShare coefficients, replace the Docking Station System Board.
  • R
    Ryan StewartSep 23, 2025
    What to do if there is Communication error between Portable and Docking Station on Masimo SET Radical-7?
    • C
      coxjoseSep 23, 2025
      To solve the communication error between the Portable and Docking Station on your Masimo SET Medical Equipment, verify the software versions of the Portable and Docking Station System Boards using the Portable menu. If the versions are incorrect or not displayed, reload System Application Software for that device utilizing procedures outlined in the Download Procedure. If the issue continues, replace the Interface Cable, the Portable System Board, or the Docking Station System Board.
  • E
    Eric BurnsSep 9, 2025
    What to do if Masimo SET Radical-7 has Display Backlight Will Not Illuminate?
    • T
      Timothy HansonSep 9, 2025
      If the display backlight will not illuminate on your Masimo SET Medical Equipment, replace the Portable System Board, or replace the Display.
  • D
    Diane GarciaSep 12, 2025
    What to do if Masimo SET Radical-7 experiences a Docking Station Watchdog Failure?
    • A
      Amanda RossSep 12, 2025
      If your Masimo SET Medical Equipment experiences a Docking Station Watchdog Failure, replace the Docking Station System Board, or replace the Portable System Board.
  • J
    Joseph VasquezSep 3, 2025
    What to do if the Masimo SET Medical Equipment has a NiMH Battery Fault?
    • A
      agarciaSep 3, 2025
      If your Masimo SET Medical Equipment indicates a NiMH Battery Fault, replace the NiMH battery. If the problem persists, replace the Portable System Board.
  • G
    Gerald CraigSep 1, 2025
    What to do if the Masimo SET Radical-7 display does not rotate manually?
    • Y
      Yvonne BarberSep 1, 2025
      If the display on your Masimo SET Medical Equipment does not rotate manually, check for flex interference and replace the Swivel Connector if needed. If the issue continues, replace the Keypad or the Portable System Board.

Summary

Overview of Radical-7 Service Manual

About this Manual

This section describes the purpose and scope of the service manual for the Radical-7 Pulse CO-Oximeter.

Warnings, Cautions, and Notes

Explains the meaning of warning, caution, and note labels used throughout the manual to ensure user safety and device integrity.

Maintenance Procedures

Introduction to Maintenance

This chapter covers testing, cleaning, battery, and fuse replacement procedures for the Radical-7 pulse oximeter.

Cleaning the Radical-7 Oximeter

Provides detailed instructions for cleaning the display panel and outer surfaces of the Radical-7 oximeter.

Battery Operation and Maintenance

Battery Operation Overview

Introduces the battery types for the Radical-7 Handheld and Docking Station and the initial charging requirements.

Battery Charging Process

Details the procedure for charging the handheld and docking station batteries, including LED indicators and charging times.

Replacing the Handheld Battery

Step-by-step guide on how to safely remove and replace the battery in the Radical-7 Handheld unit.

Replacing the Docking Station Battery

Instructions for replacing the battery within the Radical-7 Docking Station, including necessary precautions.

Fuse Replacement Procedure

Provides instructions on how to safely replace fuses in the Radical-7 unit, including fuse specifications and precautions.

Software Management

Software Overview

Explains the Radical-7's software configuration, board independence, and compatibility requirements for upgrades.

Software Download and Update Procedure

Detailed steps for downloading and installing software updates onto the Radical-7 Handheld and Docking Station systems.

Performance Verification Tests

Performance Verification Overview

Outlines the procedures for testing the Radical-7 pulse oximeter's performance after repairs or during routine maintenance.

Power-On Self-Test Verification

Describes the steps to verify the Power-On Self-Test functionality, including LED indicators and audible signals.

Key Press Button Functionality Test

Details the process for verifying that each soft key button on the Radical-7 acknowledges presses with beeps or display changes.

Alarm Limit Test Procedure

Instructions for testing the High and Low Saturation and Pulse Rate alarm limits by adjusting parameters and verifying display.

Monochrome Display Contrast Test

Procedure to test and adjust the display contrast settings on monochrome Radical-7 units.

Testing with Masimo SET Tester

Guides the user on using a Masimo SET Tester to verify SpO2 and pulse rate measurements and alarm functions.

%SpCO (Carboxyhemoglobin) Testing

Details the procedure for performing a functional test of the %SpCO measurement on the Radical-7.

%SpMET (Methemoglobin) Testing

Details the procedure for performing a functional test of the %SpMET measurement on the Radical-7.

PVI (Pleth Variability Index) Testing

Describes how to perform a functional test for the PVI feature, noting that values fluctuate.

Nurse Call Test Procedure

Provides steps to test the Nurse Call functionality by measuring resistance under different alarm conditions.

Analog Output Signal Test

Instructions for testing the Analog Output signals for SpO2 and Pulse Rate using a multimeter.

Serial Port Communication Test

Details how to validate the serial port communication using a PC or printer and specifies communication parameters.

Battery Status and Charging Test

Procedure to verify the charging status and functionality of both the handheld and docking station batteries.

Repair Procedures and Safety

Repair Safety Precautions

Crucial safety warnings and precautions to be followed before disassembling or repairing the Radical-7 unit.

General Repair Procedures

General guidelines for examining the unit, noting serial numbers, software versions, and checking warranty status before repair.

Troubleshooting Device Failures

Comprehensive troubleshooting guide with symptoms, causes, and corrective actions for various device failures.

Repair and Return Process

Masimo Repair Policy

Masimo's policy regarding warranty repair and service, emphasizing the need for authorized personnel.

Unit Return Procedure

Detailed steps for cleaning, decontaminating, packaging, and shipping the unit for service or return.

Analog Output and Nurse Call Test Details

Docking Station Power Supply Measurement

Portable Functional Block Diagram

Docking Station Functional Block Diagram

Radical-7 Isolation Block Diagram

Portable Wiring Interconnect Diagram

Docking Station Wiring Interconnect Diagram

Radical-7 Color Model Parts List

Radical-7 Monochrome Model Parts List

Docking Station Parts List (RDS-1, RDS-1 B, RDS-3)

RDS2 Docking Station Parts List

Masimo SET Radical-7 Specifications

General IconGeneral
TechnologyMasimo SET
DisplayColor Touchscreen
SpO2 Range0-100%
Pulse Rate Range25-240 bpm
Perfusion Index (PI) Range0.02-20%
Pleth Variability Index (PVI) Range0-100%
HbCO Range0-40%
SpHb Range0-25 g/dL
SpCO Range0-40%
ORI Range0-100%
Pulse Rate Accuracy±3 bpm
Battery LifeUp to 4 hours
SpO2 Accuracy±2% (70-100%)
ConnectivityWi-Fi
Power Supply100-240 VAC, 50/60 Hz
Operating Temperature5°C to 40°C
Humidity Range10-95% (non-condensing)
AlarmsAudible and visual alarms
Measurement ParametersSpO2, Pulse Rate, PI, PVI, SpHb, SpCO, SpMet, ORI
Storage Temperature-20°C to 60°C