Sensor O
PLACEMENT ISSUE
Error Code/
System Fault
Call VEP
Device requires
servicing
Place sensor on user
Conrm emitter is on
nail or top of foot/hand
and detector is on the
more eshy area
Conrm sensor is not
wrapped too tightly
Replace sensor if not
resolved
Confirm emitter is
aligned directly over
detector
Remove from any direct
light (cover sensor)
Change the site of
sensor (eg. foot vs digit)
Conrm site of
sensor is warm
Conrm site is well
perfused and has
unrestricted blood ow
Try on parent or
caregivers nger
Masimo® Rad-G® Quick Start Guide
©2024 Ontario Ventilator Equipment Pool. All rights reserved.
Date of Issue: 2022-10-19. Rev. 2024-02-13.

Other manuals for Masimo Rad-G

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Masimo Rad-G and is the answer not in the manual?

Masimo Rad-G Specifications

General IconGeneral
TypeHandheld Pulse Oximeter
Ingress ProtectionIP22
Operating Temperature5°C to 40°C
Storage Temperature-20°C to 60°C
SpO2 Range0% to 100%
Pulse Rate Range25 – 240 bpm
Battery TypeRechargeable Lithium-ion
ConnectivityBluetooth
Perfusion Index Range0.02% to 20%
PVI Range0% - 100%
Humidity10% to 95% (non-condensing)