EasyManua.ls Logo

Max-Air Systems CAPR - 19 Instructions for Use: 2270-04 HE DLC Hood with Heavy Loading Filter (HLF)

Default Icon
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
P/N 03521015 Rev K Page 71
19. Instructions For Use: 2270-04 HE DLC Hood with Heavy
Loading Filter (HLF)
Intended Use
The MAXAIR
®
CAPR
®
Hood Systems are intended to lter
aerosolized and droplet particulates and provide high level uid
resistance.
Hoods are designed for single use applications.
For Use With
The MAXAIR 2270-04 HE DLC Hood is a component of the
MAXAIR CAPR Hood Systems and provides optimum performance
when donned with a MAXAIR CAPR Powered Air Purifying Respirator
(PAPR) 2075-03 Helmet (optionally with 2076-03 Helmet w/chin bar).
CAPR 2270-04 HE DLC Hood
Standard System Conguration
1. 2075-03 Helmet w/ Power Cord 4. 2270-04 HE DLC Hood
2. 2500-36TSC Battery 5. 2170-26 Heavy Loading
Filter
3. 2000-76 Battery Belt 6. 2600-01 Battery Charger
Materials
Hood Lens is polycarbonate
Hood Filter is polypropylene
Hood Shroud is polypropylene coating on polypropylene scrim
Heavy Loading Filter is polypropylene/modacrylic blend
Specications
Temperature Limits: 49
0
C
APF = 25
Regulatory
All areas except lter pass ASTM F1671
Filter passes AATCC TM 127 @>50 cm H2O
O.N. 2270-04SM
Small-Medium
PN 07831036
O.N. 2270-04ML
Medium-Large
PN 07831037
WARNING
Use only if package is received unopened and contents are
undamaged. If damage is noted, contact the shipper for replacement
or repair.
Prior to using any MAXAIR
®
System or component, be sure to be
familiar with the system’s NIOSH approved conguration.
DO NOT use if any component is damaged. If any components are
damaged or contaminated and therefore unt for safe and effective
use, they should be replaced immediately.
Only trained and experienced personnel who have read and under-
stand the User’s Instructions should use this product.
Failure to follow the User’s Instructions, these and in the MAXAIR
Systems User’s Instructions, P/N 03521015, may be hazardous to
the user’s health.
The institution using this product in any application is responsible for
determining the appropriateness of this equipment relative to regula-
tory requirements. Bio-Medical Devices Intl, Inc. does not recom-
mend the appropriate systems for a particular institution or facility.
Use only MAXAIR Systems/ NIOSH approved compatible
components.
NOT for use in atmospheres immediately dangerous to life or health
(IDLH), and atmospheres containing less than 19.5% oxygen, or
more than 25% oxygen.
Flammability Level I: fabric may burn if exposed to open ame.
Follow current local regulations governing biohazard waste to safely
dispose of used shrouds.
If you need more information, contact your BMDI Sales
Representative, or call BMDI customer service at 1-800-443-3842.
!
Symbol Denitions
Warning, Caution, or Note Fluid Resistance
Order Number Part Number
!
O.N. PN
1
2
4
5
6
3

Table of Contents