WebDAQ 504 User's Guide Specifications
21
Parameter Condition Specification
Crosstalk
1 kHz
110 dB
CMRR f
in
≤ 1 kHz
40 dB min
47 dB typ
SFDR
f
in
= 1 kHz, 60 dBFS
120 dB
Input impedance
Differential 305
AI(shield) to chassis ground
50 Ω
Note 1: If you are using an IEPE sensor, use this equation to make sure your configuration meets the IEPE compliance
voltage range
(V
commo n-mode
+ V
bias
± V
full-scale
) must be 0 to 19, where
V
commo n-mode
is the common-mode voltage applied to the WebDAQ 504.
V
bias
is the bias voltage of the IEPE sensor.
V
full-scale
is the full-scale voltage of the IEPE sensor.
Table 2. Measurement accuracy specifications
Measurement Conditions
Percent of Reading
(Gain Error)
Percent of Range
(Offset Error) Note 2
Calibrated
Maximum
(0 °C to 50 °C)
±0.34% (±0.03 dB)
±0.14% (±7.1 mV)
Typical
(25 °C ±5 °C
±0.07% (±0.006 dB)
±0.01% (±0.5mV)
Note 2: Range = 5.1 V
pk
.
Table 3. Idle channel and noise density specifications
Idle Channel 51.2 kS/s 25.6 kS/s 2.048 kS/s
Noise
97 dBFS
99 dBFS
103 dBFS
50 µV
rms
40 µV
rms
25 µV
rms
Noise density
310 nV/√Hz
350 nV/√Hz
780 nV/√Hz
Table 4. Total harmonic distortion (THD) specifications
Input Amplitude Condition Specification
Input amplitude
1 dBFS
1 kHz: 95 dB
8 kHz: 87 dB
20 dBFS
1 kHz: 95 dB
8 kHz: 80 dB
Intermodulation distortion
(–1 dBFS)
DIN 250 Hz/8 kHz
4:1 amplitude ratio
80 dB
CCIF 11 kHz/12 kHz 1:1 amplitude
ratio
93 dB
Digital input/output
Table 5. Digital input/output specifications
Parameter Specification
Digital type CMOS (Schmitt trigger) input / open drain output
Number of I/O
One port of 4 bits
Configuration
Each bit can be independently configured for input or output
Power on conditions
Power on reset is input mode
Pull-up configuration
Each bit is pulled up to 5 V with a 100 kΩ resistor
Input frequency range DC 10 kHz (Note 3)
Input high voltage threshold
1.9 V min, 3.6 V max
Input low voltage threshold
2.3 V max, 1.0 V min
Schmitt trigger hysteresis
0.6 V min, 1.7 V max

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Measurement Computing WebDAQ 504 and is the answer not in the manual?

Measurement Computing WebDAQ 504 Specifications

General IconGeneral
BrandMeasurement Computing
ModelWebDAQ 504
CategoryData Loggers
LanguageEnglish

Related product manuals