COMPACT PIEZO LED
WARNING: Infectionscontrol.
To ensure maximum safety of the patient and
the operator, before using all the reusable
the instructions of the
17
.
CAUTION: Contraindications. Allow
reusable, autoclavable items (the handpiece,
gradually return to room temperature after
cooling process must not be accelerated.
WARNING: Breakageandwear-outof
the insert tips.
Deformed or otherwise damaged insert tips
These insert tips must never be used.
fragments remain in the treated part and,
remove them.
The patient must be instructed to breathe
through his nose during the treatment, or
a dental dam must be used to prevent the
insert tips.
its integrity before and during each use. If
there is a decrease in performance, replace
it. The level of wear of the most common
correctly:
•
printed on the card has a red line
indicating the wear limit;
• If the insert tip has a length less than
the wear limit, its performance will be
of a new one and replacement is
recommended.
If the layer of titanium nitride (golden
insert tip must be replaced. The use of a worn
the insert tip damages it and is therefore
worn out.
the insert tip is intact, especially in its apical
part.
During the intervention, avoid prolonged
contact with retractors or with metallic
instrumentation in use. Do not exert excessive
pressure on the insert tips during their use.
WARNING: OnlyuseoriginalMECTRON
inserttips,accessories,andspareparts.
CAUTION:
WARNING: Prior to any use, system
components must be inspected for damage.
Do not use if damage is apparent.
WARNING: Do not operate the device if
Immediately replace the handpiece.
NOTE: In case of accident attributable to the
device during correct use and in accordance
with the intended use, a report must be
made to the Competent Authority and
manufacturer indicated on the product label.