6.3 DryingandLubrication
6.3.1 NecessaryMaterials
• Compressed air;
•
• Medical grade lubricant.
NOTE:
Spray or similar.
Thoroughly dry all parts of the scaler
handpiece, scaler nose cone, and light
pipe by blowing them with compressed
air;
1
CAUTION: The scaler handpiece
electrical contacts must be dry at the
start and end of the sterilisation cycle,
before the cord is connected to the
electrical contacts of the connector
are perfectly dry, if necessary dry them
blowing with compressed air.
CAUTION: Before starting the
the insert tip is thoroughly dry both
internally and externally. To do this,
blow compressed air both externally and
through the internal passage hole. This
will prevent the appearance of stains,
the insert tip.