14
COMPACT PIEZO LED
At the end of treatment stow the scaler
handpiece on its housing.
8
4.5 ImportantInformationontheInsertTips
WARNING:
• Diamondcoatedinserttipsare
SINGLEUSE.The diamond coated
insert tips are intended to be used on
an individual patient during a single
surgical procedure and then discarded.
The diamond coated insert tips cannot
therefore be reprocessed since they
cannot be cleaned properly. Soil
residues might remain adhered to the
diamond coating even after cleaning and
of another patient.
•
visibly worn out, the insert tip must be
replaced. Use of an overly worn out
• Do not activate the handpiece while the
insert tip is in contact with the treated
part, so that the electronic circuit can
of the insert tip and allow its optimal
performance.
• Always inspect the insert before and
during the operation for any damage.
If a damage or drop in performance
is noted replace it with a new one.
Insert showing signs of deformation or
• Use original Mectron insert tips only. Use
to voiding the warranty, damages the
being able to screw the original insert
Moreover, the device settings are tested
and guaranteed to operate correctly
only when original Mectron insert tips
are used.
• Do not change the shape of the insert
•
any type of deformation.
• Do not attempt to sharpen the insert tip
used.
•
the insert tip and of the handpiece are
perfectly clean – see
17
.
•
exert excessive pressure on the inserts
during use. Apply a light force on the
• The Mectron inserts vibrate with a
longitudinal oscillation, with forward and
the tooth surface. Move the handpiece
lateral pressure.
• Do not aim the instrument directly on
the surface of the enamel or implant.
tangentially to the surface of the tooth
or implant.
•
instrument or leverage the operative
part. The inserts must be left free to
vibrate.