COMPACT PIEZO LED
6.4.2 SterilizationMethod
WARNING: Infectioncontrol-
Sterilisableparts.Thoroughly remove all
residual organic matter before sterilisation.
WARNING: Diamond coated inserts are
SINGLEUSEONLY. The diamond coated
inserts are intended to be used on an
individual patient during a single surgical
procedure and then discarded. The diamond


CAUTION: DONOT
The console will no longer function.
CAUTION: DONOT

CAUTION
instruments in one autoclave cycle, ensure
that the autoclave maximum load is not
exceeded.
WARNING: Riskofcontamination.DO
NOT use a gravity displacement autoclave to

intruments. The operating cycle of a gravity
displacement autoclave DONOT ensure

recessed features of the instruments.
WARNING


CAUTION: Never use any other

incompatibility with the materials used in the
construction of the components.
CAUTION: DONOTUSE the following







CAUTION: DONOT
instruments by using:
Hydrogen peroxide;
Peracetic acid system;



The scaler handpiece and other sterilisable
accessories are manufactured with materials
that resist a maximum temperature of 135 °C
for a maximum time of 20 minutes.
CAUTION: DONOTexceed 135 °C (275

Once the scaler handpiece and the other
sterilisable accessories have been put
into bags individually, perform the steam
sterilisation process in the autoclave.
CAUTION: Do not sterilise the scaler
handpiece with the insert screwed in.
CAUTION:

sterilisation pouch compliant with standard

The sterilisation process validated by Mectron

a SAL 10

by setting the parameters indicated
below:
Typeofcycle

Minimumsterilisationtemperature:

Minimumsterilisationtime:
4 minutes.
Minimumdryingtime: 20 minutes.
All stages of sterilisation must be carried out
by the operator in compliance with the most
current version of standards: UNI EN ISO
      
