08.0 Replacement of the fuses
DANGER : Switch off the apparatus.
Always turn off the apparatus by means of the switch (Fig.5 - Ref.F) and disconnect it from the
power outlet before carrying out the following maintenance activities.
1 Insert the at tip of a screwdriver into the recess in the fuse compartment below the power
socket and use it as a lever (Fig.12 - Ref.A).
2 Pull out the fuse compartment (Fig.12 - Ref.B).
3
DANGER: Replace the fuses, using fuses of the type indicated on the identication label on
the bottom of the apparatus. Depending on the mains power supply, these fuses may be rated
as follows:
230 VAC 500 mA T
115 VAC 1 mA T
100 VAC 1,2 mA T
4 Put the compartment back into place (Fig.12 - Ref.B).
Fig. 12
A B
09.0 Disposal procedures and precautions
- This equipment must be disposed of and treated as waste requiring separate collection;
- At the end of the life-cycle of this equipment, the purchaser is entitled to return the equipment
to the dealer supplying new equipment. Instructions for disposal are available from Mectron;
- Failure to comply with the foregoing points may entail punishment in accordance with Directive
2002/96/EC.
DANGER: Hospital waste.
Treat the following items as hospital waste:
- Inserts, when worn or broken;
- Irrigator, after each treatment;
- Tube of the peristaltic pump, after 8 sterilising cycles;
- Torque wrench for tightening inserts, when worn or broken.
06.8 Autoclave sterilising of the tting between the cord and the tube of
the peristaltic pump
1 Prepare a solution of enzymatic detergent with a pH of between 6 and 9, following the
manufacturer’s instructions.
WARNING: Carry out the CLEAN function on the tting of the peristaltic pump tube (Point
06.1).
2 Clean the tting of the peristaltic pump tube with the enzymatic solution prepared as above.
3 Rinse with distilled water to eliminate any residues of detergent and dry thoroughly.
NOTE: Do not use hydrogen peroxide to disinfect the tting. Use only disinfectants with a
neutral pH. Always rinse with sterile water.
4 Seal the tting individually in a a disposable bag.
5 Sterilise the tting in a steam autoclave.
07.0 Regular maintenance
07.1 Shelf Storage
If the device is not used for several days running, follow the recommendations indicated below:
1 Carry out the cleaning cycles for the water circuits using the CLEAN function (see point “06.1”).
2 Disconnect the device from the power mains.
3 Carry out cleaning, disinfection and sterilisation (see Section “06.0”)
07.2 Power-supply cable
DANGER: Check regularly that the power cable is intact. If it is damaged, replace it with an
original Mectron spare.