05.5 - Protection Systems and Alarms
The device has a diagnostics circuit that is used to recognize tripping of the protection
systems and of the alarms. These are shown on the display, as follows:
ERR1 Signal that the general protection system has been activated:
- Piezosurgery handpiece not connected to the device.
- Tuning circuit not working correctly.
- Handpiece failure.
ERR 2 Signal that the general protection system has been activated:
- Piezosurgery handpiece not connected to the device.
- Tuning circuit not working properly.
- Handpiece failure.
ERR 3 For the manufacturer only - Contact the Mectron Service Centre.
ERR 4 Signal that the general protection system has been activated:
- Tuning circuit not working properly.
- Handpiece failure.
ERR 5 Signal that the tuning scan has failed:
- Insert not correctly secured to the handpiece.
- Insert worn, broken or deformed.
- Electrical contacts of cord wet
ERR 6 Signal that the “power supply out of range” protection system has been tripped.
ERR 8 Signal indicating an error in communications with the scaler module.
ERR 9 Signal indicating an error in pump operation:
- Check whether the pump and tube are correctly installed.
- Check whether the pump is prevented from turning
ERR CKS Error signal. Firmware. Firmware checksum incorrect:
- If the equipment is working properly, call as soon as possible.
- If the equipment is not working correctly, stop using it and call the Mectron
Service Centre immediately.