MANUAL OF USE AND MAINTENANCE
13.0 - TECHNICAL DATA
Device in accordance with Directive 93/42/EEC: Class II a.
Class according to EN 60601-1: I
Type B
IP 20 (device)
IP X8 (footswitch)
Device for intermittent operation: 60” ON 30” OFF.
Power-supply voltage: 230 VAC ± 10 % 50/60 Hz.
115 VAC ± 10 % 50/60 Hz (optional).
100 VAC ± 10 % 50/60 Hz (optional).
Max. power absorbed: 70 W.
Fuses: Type 5 X 20 mm 230 VAC 2 X 500 mA T.
Type 5 X 20 mm 115 VAC 2 X 1 A T.
Type 5 X 20 mm 100 VAC 2 X 1,2 A T.
Working frequency: Automatic scan.
From 24 to 29 KHz.
Average power applied to the handpiece:
Adjustable to 6 power levels:
ROOT mode Power: ENDO: 2,8 W
Power: PERIO: 4,6 W
BONE mode Power: QUALITY1: 16 W
Power: QUALITY2: 16 W
Power: QUALITY3: 16 W
Power: SPECIAL: 4,6 W
Delivery rate of the peristaltic pump: Adjustable via the keypad.
In the ROOT power mode: 6 delivery rates:
From 0 to approx. 100 ml / min.
In the BONE power mode: 5 delivery rates:
From 15 to approx. 100 ml / min circa.
Protection systems and tripping time
of the APC: No handpiece.
Tripping time 20 ms.
Discontinuity of cord lead.
Tripping time 20 ms.
Insert not correctly tightened or broken.
Tripping time 80 ms.