22 / 32 Piezosurgery // GB
MANUAL OF USE AND MAINTENANCE
06.7 - Autoclave sterilisation of the peristaltic pump tube
1 Clean the peristaltic pump tube.
2 Disinfect with a mild disinfectant solution having a neutral pH and dry it thoroughly.
3 Seal the tube inside an individual disposable bag.
4 Autoclave sterilise the tube.
06.8 - Autoclave sterilisation of the connection between cord and
peristaltic pump tube connection
1 Clean the connection between cord and peristaltic pump tube connection.
2 Disinfect with a mild disinfectant solution having a neutral pH and dry it thoroughly.
3 Seal the connection inside an individual disposable bag.
4 Autoclave sterilise the connection.