MANUAL OF USE AND MAINTENANCE
05.6 - Instructions for Use
1 Open the air intake on the drip system.
2 Fit the required insert (Fig.10 - Ref.A) onto the Piezosurgery handpiece and lock it into
place using one of the wrenches included in the supply.
3 To use the torque wrench correctly (Fig.10 - Ref.B), proceed as follows:
- screw the insert onto the handpiece as far as it will go
- hold the wrench and the casing of the handpiece very firmly
- turn the wrench in a clockwise direction until the clutch engages (the outer casing of
the wrench slides against the casing of the handpiece). The insert is now locked prop-
erly into place.
4 Make sure that the Piezosurgery handpiece is correctly connected to the handpiece
connector (Fig.5 - Ref.D).
5 Check the display to see the type of power that has been set (See point 05.3). If the
type of power required differs from the type that has been set, use the key for select-
ing the type of power (Fig.6 - Ref.D). This can be pressed to toggle between the
ROOT and BONE functions.
6 Check the display to see the power level that has been set (See point 05.3). If the type
of power required differs from the level that has been set, use the key for selecting
the power level (Fig.6 - Ref.C). This can be used to choose between the following
options, depending on the type of power that has been set:
- ROOT ENDO - PERIO
- BONE QUALITY 1 - QUALITY 2 - QUALITY 3 - SPECIAL
7 Check the display to see the delivery rate of the peristaltic pump (see point 05.3). If the
delivery rate required is other than the level that has been set, use the PUMP + and
keys (Fig.6 - Ref. B) to choose between the following, depending on the type of power
that has been set:
- ROOT 6 delivery rate levels From 0 to 5
- BONE 5 delivery rate levels From 1 to 5
NOTE: Sprayless treatment is only possible in the ROOT function, setting the delivery
rate of the peristaltic pump PUMP to nought (Fig.7 - Ref.A). No notches should be
shown on the display.
05.7 - Rules for Keeping the Device in Proper Working Order
1 Check the state of wear of the inserts periodically and replace any for which a drop in
performance is noted.
2 Do not alter the shape of the inserts by bending or filing them.
3 Replace any insert that has become deformed or damaged by impacts.