Piezosurgery // GB 27 / 32
MANUAL OF USE AND MAINTENANCE
PROBLEM POSSIBLE CAUSE SOLUTION
Cord not connected to the
device
Lack of continuity of a lead
in the cord.
Handpiece failure.
Malfunctioning of the tuning
circuit.
Connect the cord to the de-
vice
Contact the nearest author-
ised MECTRON Service
Centre.
Contact the nearest author-
ised MECTRON Service
Centre.
Contact the nearest author-
ised MECTRON Service
Centre.
No liquid comes out of the
insert during operation.
The insert is of the type with
no through-flow of liquid.
The PUMP function is set to
nought.
The bag of liquid is empty.
The drip system air intake
has not been opened.
The tubes of the drip system
and of the pump have not
been correctly installed.
The insert is clogged.
The handpiece is clogged.
The device is switched on
but will not work. The mes-
sage ERR appears on the
display.
Use an insert of the type with
through-flow of liquid.
Press the + key on the front
panel of the device for the
PUMP function.
Replace the bag with a full
one.
Open the air intake of the
drip system.
Check the connections of
the tubes.
Free the passage in the in-
sert through which the wa-
ter passes.
Contact the nearest author-
ised MECTRON Service
Centre.