26
WARNING: Do not press the foot pedal


the foot pedal.
WARNING
only be used in place of the foot pedal

the operation even if the foot pedal is not

4.14 SafetyRequirementsBeforeandDuringUse
WARNING: Before starting the activity,
always assure having a set of spare parts

the event of a failure or problem.
WARNING: Useofnon-original
Piezosurgery/Mectroninserts: this can
completely damage the threading of the

and causing injury to the patient.
CAUTION:Contraindications. Do not
operate on metal or ceramic prosthetic

cause them to detach.
WARNING: Contraindications. Do not
use the PIEZOSURGERplus on patients


operator.
WARNING: Checktheconditionof
thedevicebeforeoperating.
sure that there is no water under the device.


accessories, as well. In case of problems or



Service.
WARNING: Infectioncontrol.Firstuse:
all reusable parts and accessories (whether



the use, in accordance with the instructions

Manual. Subsequentuses: after each
treatment, clean and sterilise all reusable
parts and accessories, according to the

Sterilisation Manual.
CAUTION:Contraindications. After
having sterilised the handpiece, the torque
wrench and any other sterilisable accessories
in the autoclave, wait for them to cool
completely before reusing.
WARNING: Only use original

parts.
CAUTION:Theelectricalcontactsinside
thecordconnectormustbedry. Before
connecting the handpiece to the device,

connector are perfectly dry, in particular,
after the autoclave sterilisation cycle. Dry
the contacts, if necessary, by blowing
compressed air onto them.
WARNING
handpiece cooling, always activate it with the


pump button.
WARNING
functionality before and during use. Assure


or the pump is faulty.