42
InformationMessage
of an anomaly; the device continues to
operate, but with limited functionalities.
current message and the indication of
the total number of anomalies detected.
3
ErrorMessage ActionsRequired Description
MED-001
Replacethehandpiece.
Replace the handpiece. Malfunction of the tuning
circuit or faulty handpiece.
MED-002
Replacethehandpiece.
Replace the handpiece. Malfunction of the tuning
circuit or faulty handpiece.
MED-004
Please check that the
handpieceisusedproperly.
If the problem persists,
replaceit.
handpiece is used properly.
If the problem persists,
replace it.
Malfunction of the tuning
circuit or faulty handpiece.
MED-005
Tighten the insert tip.
If the problem persists,
replacethehandpiece.
1. Insert not properly fastened
onto the handpiece.
2. If the problem persists,
replace the handpiece.
Drop of the handpiece
electrical insulation.
3. If the problem persists,
replace the insert.
deformed.
4. If the problem persists,
dry the electrical
contacts by blowing
compressed air.
In rare cases, the presence
of humidity in the electrical
contacts of the cord may
occur.
MED-006
Internalproblem.
“SHUTDOWN”thedevice
thenstartitagain.
then start it again.
protection.
MED-016
Channeloutofuse.
channel module failed.
MED-017
Makesurethatthepump
andtubearecorrectly
installed
and tube are correctly
installed.
Failure in pump functionality.
rotation.