MEDIVATORS REPROCESSING SYSTEMS
21
Caution: Verify fluid exiting the open end of the endoscope or tubing is not facing a direction
that would allow fluid to be sprayed on personnel, or electrical fixtures. The endoscope should
always be fully immersed under water.
Note: Detergent and rinse water cleaning solutions can be drawn from either a clean
container for a single-pass through Scope Buddy, or recirculated from the sink through Scope
Buddy. Scope Buddy can be used in either format depending on the facility’s endoscope
cleaning procedure.
Figure 6 Universal Cleaning Adapter hook-up with 3-way connector
Flushing an Endoscope with Detergent:
1. Plug the unit in and then turn it on by pressing the power switch located on the side of the unit. The
timer display will now illuminate. See Figure 7.
Sco
e Budd
Unit
Universal Cleaning Adapter
with 3-way Connector
Fluid Intake Line